ECG calibration
Also calledstandardization · calibration pulse · cal mark · standardization mark · calibration signal
Definition
The reference settings that make ECG amplitudes and intervals comparable between recordings. The standard gain is 10 mm/mV: a 1 mV signal moves the trace 10 mm (two large boxes), so each 1 mm small box equals 0.1 mV. The standard paper speed is 25 mm/s: each 1 mm small box equals 0.04 s and each 5 mm large box 0.20 s. A calibration signal on the tracing lets the reader confirm the gain; on analog machines the 1 mV (10 mm) pattern is checked on each tracing, while digital machines verify it automatically.